CertikLabs · Certificate of Analysiscertiklabs.com/v/CL-7F92-A19D
Sample report

CL-7F92-A19D

Semaglutide · 5 mg · Lot L-2401-77

Verified · Pass
All assays within spec · Anchored Apr 28, 2026
Overall confidenceHigh · 99%
Identity
High
MatchPass
Target: Semaglutide (C₁₈₇H₂₉₁N₄₅O₅₉)

LC-MS [M+4H]⁴⁺ at m/z 1029.6 confirms target peptide. No isoform contamination detected.

Purity (HPLC)
High
99.4%Pass
Spec: ≥ 98.0% · USP <621>
96spec 98100100

1.4 pp above minimum. Single dominant peak; impurity sum 0.6%, all individual <0.3%.

Concentration
High
100.6%Pass
Spec: 95.0–105.0% of label · LC-MS
90spec 95105110

Within ±5% acceptance window. ±0.4% analytical uncertainty (n=3, RSD 0.4%).

Endotoxin (LAL)
< 0.5 EU/mg
Limit < 5 EU/mg
Water content (KF)
3.2%
Spec ≤ 6.0%
Appearance
White lyophilate
Conforms
What this means. This sample is the substance it claims to be (identity), is at least 99% pure (purity), and contains the labeled amount within ±5% (concentration). All three primary acceptance criteria are met with high analytical confidence (RSD < 0.5%, n=3). No reanalysis required.
Method · HPLC-UV
ColumnC18, 4.6 × 150 mm, 3.5 µm
Mobile phase0.1% TFA in water / acetonitrile gradient
Flow1.0 mL/min
DetectionUV 214 nm
Injection10 µL · 1 mg/mL
Method · LC-MS
SourceESI+
Rangem/z 200–2200
CalibrationExternal, ESI tuning mix
Result[M+4H]⁴⁺ confirms target
Content100.6 ± 0.4% of label
Endotoxin · LAL
MethodKinetic chromogenic
Limit< 5 EU/mg (USP)
Result< 0.5 EU/mg (PASS)
Sample chain of custody
ReceivedApr 24, 2026 · 09:14 UTC
LabAurum Bio · ISO 17025 #ABL-2241
AnalystC. Mendes
Reviewed byDr. R. Halberg
Transparency & data sources
Every result is auditable

Each headline result links to the originating method, instrument run, raw data file, standard operating procedure, and reference standard used for comparison. Files are content-addressed (SHA-256) and anchored on-chain — tampering breaks the link.

Identity
LC-MS · ESI+ deconvolution
Instrument
Waters Xevo G2-XS QTof
Source
ESI+, capillary 2.5 kV
Mass range
m/z 200–2200
Reference
USP Semaglutide RS Lot R094H0
Purity (HPLC)
RP-HPLC-UV · gradient elution
Column
C18, 4.6 × 150 mm, 3.5 µm
Mobile phase
0.1% TFA / MeCN gradient
Detection
UV 214 nm
Compendial
USP <621> system suitability met
Concentration
LC-MS quantitation · external std
Standard curve
5-point, r² = 0.9998
Replicates
n = 3, RSD 0.4%
Acceptance
95.0–105.0% of label
Reference
USP Semaglutide RS Lot R094H0
All linked artifacts are signed by lab key ABL-2241-KEY-04 and anchored together.
Signature & anchor
Document SHA-256
9a31f8b14c0d…b4f2c0e7
Anchor tx
0x55ce10aa…917b · Apr 28, 2026
Signed by
CertikLabs · CL-SIGN-2026-04
Re-verify
certiklabs.com/anchor/CL-7F92-A19D
← Back to verifier